Abstract

Position-sensitive photomultiplier tubes (PSPMTs) in array are used as gamma ray position detector. Each PMT converts the light of wide spectrum range (100 nm ~ 2500 nm) to electrical signal with amplification. Because detection system size is determined by the number of output channels in the PSPMTs, resistive network has been used for reducing the number of output channels. The photo-generated current is distributed to the four output current pulses according to a ratio by resistance values of resistive network. The detected positions are estimated by the peak value of the distributed current pulses. However, due to parasitic capacitance of PSPMTs in parallel with resistor in the resistive network, the time constants should be considered. When the duration of current pulse is not long enough, peak value of distributed pulses is reduced and detected position error is increased. In this paper, we analyzed the detected position error in the resistive network and variation of time constant according to the input position of the PSPMTs.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.