Abstract

A gas chromatography coupled to triple-quadrupole mass spectrometry (GC-(QqQ)MS/MS) including a soft-ionization through an atmospheric pressure chemical ionization (APCI) source based method was compared with the high resolution mass spectrometry (HRMS) standard reference method EN1948, for the analysis of polychlorinated dibenzo-p-dioxins and furans (PCDD/Fs) in stack gas emissions. The stack emission samples were collected, both, by manual method sampling (from 6 to 8h) and by long-term sampling systems (sampling time of several weeks). This work presents the first comparison of GC-(QqQ)MS/MS with APCI source with the European Standard EN1948 technique for stack gas emissions. Sample concentrations ranged from 0.5 to 596pg I-TEQ/Nm3. Comparative results in all investigated samples showed relative errors that were within ±15%. These results make GC-(QqQ)MS/MS with APCI suitable for the quantitative analysis of dioxins in the studied samples and create a real alternative tool to the reference sector GC-HRMS instruments.