Abstract

ABSTRACT This study examined the dose and image quality according to the position change of a human phantom in a CT scan. This study used an MDCT 128 Slice CT Scanner instrument. An axial scan was performed with a 16 cm CTDI phantom of a human phantom, and the dose was measured using a pencil chamber meter. The phantom was scanned 10 cm above and below the isocenter and 15 cm above the right and left. The position of the phantom is indicated by C-0 in the isocenter position, S-10 in the upper 10 cm, I-10 in the lower 10 cm, R-15 in the right 15 cm, and L-15 in the left 15 cm. The test was performed 30 times using the brain CT protocol to calculate the dose and the dose width product (DWP). The acquired images were analyzed using the ImageJ program. Statistical analysis was performed using SPSS with one-way ANOVA (p < .05). The mean DWP values of the CT scanner were C-0 31.97 mGy·cm, S-10 24.52 mGy·cm, I-10 24.28 mGy· cm, R-15 17.95 mGy·cm, and L15 17.6 mGy·cm. Compared to the isocenter (C-0), the DLP values measured at each site were 23.3% for S-10, 24% for I-10, 43.8% for R-15, and 44.9% for L15. A significant difference in the one-way ANOVA statistical process was observed (p>0.05). C-0 was measured to be 7.42 HU, S-10 7.87 HU, I-10 8.4 HU, R-15 117 HU, and L-15 13.6 HU for evaluating the image quality. Compared to C-0, S-10 was 5.39%, I-10 was 13.2%, R-15 was 57.6%, and L-15 was 83.2%. The PSNR for S-10, I-10, R-15, and L-15 was 17.37, 17.5, 16.62, and 16.37 dB, respectively. A good quality image can be obtained by positioning the subject precisely in the isocenter in the axial scan, if possible, because the irradiated dose to the subject is low, which can lead to an increase in noise in image reconstruction.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call