Abstract

A liquid chromatographic (LC) method that was previously developed for penicillin G residues in animal tissues has been adapted to milk and milk products. After protein precipitation with sodium tungstate, samples are applied to a C18 solid-phase extraction cartridge, from which penicillin is eluted, derivatized with 1,2,4-triazole-mercuric chloride solution, and analyzed by isocratic liquid chromatography (LC) on a C18 column with UV detection at 325 nm. Quantitation is done with reference to penicillin V as an internal standard. Penicillin G recoveries were determined to be > 70% on standards fortified at 3-60 ppb. Accuracy approached 100% using the penicillin V internal standard. The detection limit for penicillin G residues was 3 ppb in fluid milk. Samples may be confirmed by thermospray/LC at concentrations approaching the detection limit of the UV method.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.