Abstract

A procedure for the determination of seven indicator PCBs in soils and sediments using microwave-assisted extraction (MAE) and headspace solid-phase microextraction (HS-SPME) prior to GC-MS/MS is described. Optimization of the HS-SPME was carried out for the most important parameters such as extraction time, sample volume and temperature. The adopted methodology has reduced consumption of organic solvents and analysis runtime. Under the optimized conditions, the method detection limit ranged from 0.6 to 1 ng/g when 5 g of sample was extracted, the precision on real samples ranged from 4 to 21% and the recovery from 69 to 104%. The proposed method, which included the analysis of a certified reference material in its validation procedure, can be extended to several other PCBs and used in the monitoring of soil or sediments for the presence of PCBs.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.