Abstract

The spark source mass spectrometry was applied to the analysis of 18O in solid oxides. After some trials such method was adopted successfully that a solid specimen was wound around by gold wire and a pointed gold wire was set as a counter-electrode. After the ion source chamber was evacuated, the specimen was heated for 30 minutes by the use of the heater set in the chamber in order to remove the adsorbed water. As a result of this treatment the influence of the water was erased. The quantitative analysis was performed by comparing the density of the spectrum of 18O+ with that of 16O+. The crater sizes were measured by the microscopy and the depth of sampling was estimated at the order of 5μ. The method stated in this note will be applied validly to determine the 18O distribution in the diffusion couple or the product layer formed by solid state reaction. 2 figs., 3 tables, 2 refs.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.