Abstract

A method has been developed and validated for the quantitation of midazolam, alphahydroxy-midazolam, omeprazole, and hydroxyomeprazole from one 250 microL sample of human plasma using high performance liquid chromatography coupled to tandem mass spectrometry. The method was validated for a daily working range of 0.400-100 ng/mL, with limits of detection between 2 and 15 pg/mL. The inter-assay variation was less than 15% for all analytes at four control concentrations and the samples were stable for three freeze-thaw cycles under the analysis conditions and 24 h in the post-preparative analysis matrix. This method was used to analyze samples in support of clinical studies probing the activity of the cytochrome P-450 enzyme system.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.