Abstract
An ion-pairing high-pressure liquid chromatography procedure was developed for analysis of mixtures of oligosaccharides generated by nitrous acid cleavage of heparin. Oligosaccharides were eluted from a Hi-Chrom 5S ODS (C18) column using mixtures of acetonitrile and buffers containing 40 m m ammonium phosphate and 1 m m tetrabutylammonium phosphate. Isocratic conditions were developed for optimal separation of a number of individual disaccharides and tetrasaccharides that were characterized previously ( M. J. Bienkowski and H. E. Conrad (1985) J. Biol. Chem. 260, 356–365 ). These isocratic conditions were then coupled to obtain gradient elution conditions for the ion-pairing separations of mixtures of disaccharides and mixtures of tetrasaccharides. A comparison of the elution profiles obtained in the ion-pairing chromatography procedure with profiles obtained by anion-exchange high-pressure liquid chromatography profiles showed markedly better overall resolution by the ion-pairing procedure. As a result of this improved resolution, the new procedure showed the presence of previously unidentified products in the heparin oligosaccharide mixtures.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.