Abstract

An isocratic liquid chromatographic method with pulsed electrochemical detection is described for the determination of neomycin in the presence of its impurities. The mobile phase is composed of an aqueous solution containing 35 g/l of sodium sulphate, 1 g/l of sodium 1-octanesulfonate, 14 ml/l of tetrahydrofuran (THF) and 50 ml/l of 0.2 M phosphate buffer pH 3.0. Sodium hydroxide was added post column to enhance the detection. An investigation of different reversed-phase columns indicated that the Discovery (C18 5 microm, 250 mm x 4.6 mm I.D.) column was the most suitable. The proposed method shows high efficiency, allowing the separation of the main component neomycin B from neomycin C and 15 other impurities. A central composite design was used to assess the robustness of the method. The method showed good selectivity, repeatability, linearity and sensitivity. This method was applied to analyse commercial samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.