Abstract

The commercial availability of boron-loaded organic scintillators has led to the development of neutron detectors that operate as “electronically” black, totally absorbing spectrometers. The key to the enhanced spectroscopy is the delayed capture of nearly thermalized neutrons by 10B that can occur within a few microseconds after the energy pulse from prompt proton recoils. Accurate information regarding the mean lifetime is important for correct setting of the timing logic of the detection system to obtain good neutron detection efficiency with a low chance coincidence rate. In this paper we present an analysis of the mean lifetime for neutron capture for the boron-loaded plastic BC454. Measurements of the capture time constant obtained with a 7.62 cm diameter, 10.16 cm long detector are compared with values computed with the time-dependent Monte Carlo neutron transport code MCNP. Additional analyses using MCNP examine the dependence of the mean lifetime on the boron concentration, the detector's dimensions and the incident neutron energy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.