Abstract

The use of high resolution γ-ray spectrometry in the detailed analysis of low level samples for the 238U and 232Th decay chains is discussed. A spectrometer is described, based on a “p” type intrinsic germanium well-detector contained in a cryostat manufactured from selected low background materials, and with a sample capacity of up to 10 g. Problems of sample self absorption are discussed and the calibration of the spectrometer detailed. Checks to confirm the calibration are described and illustrative analyses of samples of interest to thermoluminescence dating given. Typical overall uncertainties are between 1 and 5 Bq kg −1, depending on the radionuclide concerned.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.