Abstract

Summary A sensitive and selective liquid chromatographic tandem mass spectrometric (LC-MS-MS) method for analysis of lovastatin in human plasma has been developed and validated. Ethyl acetate extraction was used for plasma sample preparation and simvastatin was used as internal standard. Chromatographic separation was achieved on a C18 column by isocratic elution with 83:17:0.1 (v/v) methanol-2 μM aqueous sodium acetate-formic acid as mobile phase, delivered at 1.0 mL min−1. MS-MS detection was performed using positive electrospray ionization and multiple-reaction monitoring with argon for collision-induced dissociation. Calibration plots were generated over the concentration range 0.05 to 20 ng mL−1 (r > 0.999) with a lower limit of quantification (LLOQ) of 0.05 ng mL−1. Intra and inter-day precision and accuracy were determined at four different concentrations, 0.05, 0.5, 2.0, and 10.0 ng mL−1, and precision ranged from 0.4 to 11.4% with the deviation always less than 15% (n = 5). Extraction recoveries ...

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.