This paper describes a method for speciation of Hg associated with airborne particulate matter. This method uses a mini-sampler for sample collection and analysis, thermal desorption for separating Hg species, and inductively coupled plasma mass spectrometry (ICP-MS) for identification and quantification of Hg. Coal fly ash spiked with different Hg compounds (e.g. Hg0, HgCl2, HgO, and HgS) was used for qualitative calibration. A standard reference material with a certified value for Hg concentration was used to evaluate the method. When the temperature of the furnace was programmed at a linear rate of increase of 50 degrees min(-1), different Hg compounds could clearly be separated. Three airborne particulate matter samples were collected in parallel in Toronto, ON, Canada and analyzed using this method. Reproducible results were obtained and Hg0, HgCl2, HgO, and HgS species from these samples were detected.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.