Abstract

“Headspace” technique is one of the methods for the onboard measurement of hydrogen (H2) and methane (CH4) in deep seawater. Based on the principle of an automatic headspace commercial sampler, a specific device has been developed to automatically inject gas samples from 300ml syringes (gas phase in equilibrium with seawater). As valves, micro pump, oven and detector are independent, a gas chromatograph is not necessary allowing a reduction of the weight and dimensions of the analytical system. The different steps from seawater sampling to gas injection are described. Accuracy of the method is checked by a comparison with the “purge and trap” technique. The detection limit is estimated to 0.3nM for hydrogen and 0.1nM for methane which is close to the background value in deep seawater. It is also shown that this system can be used to analyze other gases such as Nitrogen (N2), carbon monoxide (CO), carbon dioxide (CO2) and light hydrocarbons.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.