Abstract

In this paper, a highly sensitive surface plasmon photonic crystal fiber (PCF) biosensor is reported and studied to monitor glucose concentration. The suggested design is based on a well-known large mode area (LMA) single mode PCF infiltrated by a plasmonic material. Additionally, an etching process is applied to increase the biosensor sensitivity. The numerical analysis is obtained using a full vectorial finite element method (FVFEM). The suggested biosensor based on a commercial PCF with plasmonic rod achieves sensitivity as high as 7900 nm/RIU with corresponding resolution of 1.26 × 10-5RIU-1. The analysis also reveals that the proposed biosensor has a linear performance which is needed practically. Therefore, the reported biosensor has advantages in terms of fabrication feasibility and high linear sensitivity

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.