Glimepiride (Amaryl), which is a new oral antidiabetic drug in the sulfonylurea class, was analysed by using second order derivative UV spectrophotometry. The quantification of glimepiride in dimethylformamide was performed in the wavelength range of 245-290 nm at N = 6, ?lambda = 21. The second order derivative spectra was calculated using peak to peak (lambdaDMF = 263.3-268.2 nm), peak to zero (lambdaDMF = 268.2 nm) and tangent (lambdaDMF = 263.3-271.8 nm) method for calibration curves, the linearity range of 1.00-500.00 microg ml(-1) by using the second order derivative UV spectrophotometric method. The developed method was applied to directly and easily to the analysis of the pharmaceutical tablet preparations. R.S.D. were found to be 4.18% (Amaryl tablet; 1 mg) and 2.21% (Amaryl tablet; 2 mg). The method was completely validated and proven to be rugged. The limit of quantitation and the limit of detection were found as 1.00 and 0.4 microg ml(-1), respectively. This validated derivative UV spectrophotometric method is potentially useful for a routine laboratory because of its simplicity, rapidity, sensitivity, precision and accuracy.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.