Abstract

In this research, a needle trap device (NTD) packed with nanoporous silica aerogel as a sorbent was used as a new technique for sampling and analysis of formaldehyde and acrolein compounds in aqueous and urine samples. The obtained results were compared with those of the commercial sorbent Carboxen1000. Active sampling was used and a 21-G needle was applied for extraction of gas in the sample headspace. The optimization of experimental parameters like salt addition, temperature and desorption time was done and the performance of the NTD for the extraction of the compounds was evaluated. The optimum temperature and time of desorption were 280°C and 2min, respectively. The ranges of limit of detection, limit of quantification and relative standard deviation (RSD) were 0.01-0.03μgL-1, 0.03-0.1μgL-1 and 2.8-7.3%, respectively. It was found that the NTD containing nanoporous silica aerogel had a better performance. Thus, this technique can be applied as an effective and reliable method for sampling and analysis of aldehyde compounds from different biological matrices like urine, exhalation and so on.