Abstract
An equation analogous to the Van Deemter equation in chromatography is developed to account for peak broadening in capillary zone electrophoresis (CZE). This equation applies to conditions wher the peaks are symmetrical (sample zone concentration much less than background electrolyte concentration). To identify and to quantitate the effects of different contributions to the peak width in CZE, it is first necessary to put all peak profiles on the same footing by correcting them for the velocities fo different zones and for the finite length of the detector zone compared to the sample zone. Three major contributors to the peak width are identified: (1) the injection length of the sample; (2) longtudinal diffusion that takes place during the migration time between injection and detection; and (3) analyte-wall interactions. Temperature is shown not to be a major factor in peak broadening under typical experimental conditions. The predictions of our model agree well with experimentally determined peak profiles for different analytes under a variety of conditions. New expressions for theoretical plate number and resolution in CZE are presented. It is concluded that in almost all previously reported CZE separations the peak profiles were dominated by the sample injection length. This explains why the observed peak widths in CZE have been broader than anticipated.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.