A commercially available 9.4 Tesla Fourier transform ion cyclotron resonance (FTICR) mass spectrometer was applied in the analysis of tryptic digests of protein mixtures without any separation. First, the method was demonstrated on a mixture of tryptic digests of equine cytochrome c, equine myoglobin and bovine serum albumin. The same method was then applied to human plasma from a healthy blood donor. Computer programs were employed to simplify analysis of the complex spectra. The 2745 peaks in the human plasma electrospray ionization FTICR spectrum could be reduced to 1165 isotopic clusters and 669 unique masses. Out of these, 82 masses matched tryptic fragments of serum albumin with mass measurement errors less than 10 ppm, covering 93% of the sequence. Another 16 masses were assigned to tryptic fragments of transferrin, covering 41% of the sequence on the 10 ppm mass measurement error level (14 within 2 ppm). The mass measurement errors were approximately normal distributed with a standard deviation of 1.7 ppm. This demonstrates the feasibility of combining the ultra-high mass resolving power and accuracy of FTICR mass spectrometry with automated computer analysis for investigating complex biological matrices.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.