Abstract
Edge-losses occur during reflectance measurements of pigmented maxillofacial elastomer when light is scattered within a sample beyond that part of the surface exposed to the observation system of the optical device. A custom sample-holder is presented which redirects light that would not be measured during conventional reflectance measurement back into the sample. The amount of edge-loss occurring within thin layers of maxillofacial elastomer with tan pigment on black-and-white backings was found to depend on sample thickness, the backing, the beam size used during conventional reflectance measurement, and the optical term bS = (2KS + K2)1/2. Data analysis revealed a significant interaction among these four factors. Additionally, the edge-loss occurring during the tristimulus reflectance measurement of thick samples of maxillofacial elastomer with various concentrations of tan and black pigment was found to be linearly related to bS up to a limiting value, with no additional edge-loss occurring for bS values above this limiting value. Edge-loss is an important consideration during the matching of the optical characteristics of pigmented maxillofacial material to those of human skin.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.