Abstract

The purpose of this study was to assess the relationship between forces derived from in-shoe pressure distribution and GRFs during normal gait. The relationship served to demonstrate the accuracy and reliability of the in-shoe pressure sensor. The in-shoe pressure distribution from Tekscan F-Scan system outputs vertical forces and Centre of Force (COF), while the Kistler force plate gives ground reaction forces (GRFs) in terms of F<sub>z</sub>, F<sub>x</sub> and F<sub>y</sub>, as well as vertical torque, T<sub>z</sub>. The two systems were synchronized for pressure and GRFs measurements. Data was collected from four volunteers through three trials for both left and right foot under barefoot condition with the in-shoe sensor. The forces derived from pressure distribution correlated well with the vertical GRFs, and the correlation coefficient (r<sup>2</sup>) was in the range of 0.93 to 0.99. This is a result of extended calibration, which improves pressure measurement to give better accuracy and reliability. The COF from in-shoe sensor generally matched well with the force plate COP. As for the maximum vertical torque at the forefoot during toe-off, there was no relationship with the pressure distribution. However, the maximum torque was shown to give an indication of the rotational angle of the foot.