Abstract

The Technical Report Series 398 (TRS-398), Electron Dosimetry Working Party (EWDP), and Task Group 51 (TG 51) are the most important protocols for reference dosimetry. In the case of electron beam reference dosimetry, these protocols recommend using parallel-plate ionization chambers for beams with R50 values below specific thresholds. However, recent papers suggested using cylindrical chambers for reference dosimetry of all electron beam energies. Here we compared different protocols using a cylindrical chamber with the recommendations of using a parallel-plate chamber and the TRS-398 formalism for the dosimetry of several electron beam energies. We employed electron beams with nominal energies of 4, 6, 9, 12, and 15 MeV of a Varian 2100C linear accelerator, an Exradin A12, and an Exradin P11 chamber for the analysis. The results showed differences below 3% when comparing the cylindrical chamber and alternative protocols with the parallel-plate chamber and the TRS-398 formalism for electron beams reference dosimetry. These results can bring confidence in using a cylindrical chamber for reference electron beam dosimetry, which can make the electron beam dosimetry procedure simpler and faster.