Abstract

In order to derive plural values of fluorescence lifetimes simultaneously from a multi-component sample, we formulate a mathematical method for analyzing data obtained from a frequency-multiplexed phase-modulation fluorometer (FM-PMF) using an autoregressive (AR) model. Various parameter settings necessary for performing accurate data analysis based on the AR model are studied through numerical simulations. Measurement results of fluorescence lifetimes of real samples, 10 ppm quinine sulfate in 0.1 N H(2)SO(4), 10 ppm rhodamine 6G in ethanol, and their mixture with a volume ratio of 1:1, demonstrate that the proposed method works quite well.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.