Abstract

We report on the behaviour of the dark current images of the Event Detection Intelligent Camera (EDICAM) when placed into an irradiation field of gamma rays. EDICAM is an intelligent fast framing CMOS camera operating in the visible spectral range, which is designed for the video diagnostic system of the Wendelstein 7-X (W7-X) stellarator. Monte Carlo calculations were carried out in order to estimate the expected gamma spectrum and dose for an entire year of operation in W7-X. EDICAM was irradiated in a pure gamma field in the Training Reactor of BME with a dose of approximately 23.5Gy in 1.16h. During the irradiation, numerous frame series were taken with the camera with exposure times 20μs, 50μs, 100μs, 1ms, 10ms, 100ms. EDICAM withstood the irradiation, but suffered some dynamic range degradation. The behaviour of the dark current images during irradiation is described in detail. We found that the average brightness of dark current images depends on the total ionising dose that the camera is exposed to and the dose rate as well as on the applied exposure times.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.