Abstract

In the frame of the EU Coordination Action CONRAD (coordinated network for radiation dosimetry), WP4 was dedicated to work on computational dosimetry with an action entitled 'Uncertainty assessment in computational dosimetry: an intercomparison of approaches'. Participants attempted one or more of eight problems. This paper presents the results from problems 4-8-dealing with the overall uncertainty budget estimate; a short overview of each problem is presented together with a discussion of the most significant results and conclusions. The scope of the problems discussed here are: the study of a (137)Cs calibration irradiator; the manganese bath technique; the iron sphere experiment using neutron time-of-flight technique; the energy response of a RADFET detector and finally the sensitivity and uncertainty analysis for the recoil-proton telescope discussed in the companion paper.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call