This study aimed to evaluate the reasons for cochlear implant (CI) revision surgeries in pediatric and adult groups. A total of 490 CIs were used for 423 patients between August 2005 and August 2019. Among these, patients who underwent a CI revision surgery were identified retrospectively. The reasons for CI revision surgeries were classified mainly as medical problems and device failure. These were also determined according to implant brands. Of 423 patients with a CI surgery, 27 (6.3%) underwent a CI revision surgery. The revision implant rate was 4.9% for the pediatric group and 1.4% for the adult group. Five patients underwent a CI revision surgery due to medical problems (23.8%) and 16 due to device failure (76.2%) in the pediatric group, while 3 patients (50.0%) underwent a CI revision surgery due to medical problems and 3 (50.0%) due to device failure in adults. A CI surgery is safe for patients with severe-to-profound sensorineural hearing loss, although a CI revision surgery has some potential. Device failure is the most common cause in children. It may be due to early implantation, frequent fall when children learn walking, or impaired vestibular function.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: European archives of oto-rhino-laryngology : official journal of the European Federation of Oto-Rhino-Laryngological Societies (EUFOS) : affiliated with the German Society for Oto-Rhino-Laryngology - Head and Neck Surgery
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.