Abstract

We analyzed the failure mechanism in miniature x-ray tubes with a gated carbon nanotube (CNT) field emitter. The miniature x-ray tube even with a small leakage current of field emission from the CNT emitters to the gate electrode has some distortion of the electron beam trajectory from the cathode to anode passage, causing a charging at insulating parts. The charging at the insulator spacer would lead to an arcing or coupling of the anode and gate electrodes, resulting in a very unstable operation and finally a fatal failure.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.