Abstract

A fast and precise analysis of the synthetic peptide buserelin in urine using CZE-ESI-MS method has been demonstrated. Formic acid at 50 mmol/L concentration served as background electrolyte in CZE stage and it is compatible with MS detection in positive ionization mode. Two injection modes were tested, i.e. pressure (50 mbar for 5 s) and electrokinetic injection (5 kV for 5 s), of which electrokinetic injection provided better calibration parameters. Buserelin LODs were 0.47 microg/mL in water and 0.63 microg/mL in ten times diluted urine samples using pressure injection, while they were 0.32 microg/mL in water and 0.34 microg/mL in ten times diluted urine samples using electrokinetic injection. Repeatability of buserelin migration times was below 6% (pressure injection mode) and 1% (electrokinetic injection mode). Repeatability of buserelin peak area in SIM mode (m/z=620.5+/-0.5) was less than 12% (pressure injection mode) and 5.8% (electrokinetic injection mode). In this work, no interferences were observed during the analyses of spiked urine samples.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.