Abstract
A prompt gamma activation analysis facility has been constructed on the ST1 horizontal beam port at the HANARO research reactor, KAERI in 2003. The detector system consists of a high-purity Ge detector surrounded by BGO/NaI(Tl) scintillators as an annulus type to reject the Compton scattered photons. Detection sensitivity for boron was obtained from the prompt gamma-ray spectra of boric acid, B(OH)3, containing 0.1-65 μg boron. The net peak for the calculation of the boron concentration was obtained by eliminating the sodium 472 keV peak, involved in the boron 478 keV peak. The biological samples used are NIST SRMs such as Peach Leaves, Apple Leaves, Tomato Leaves, Spinach Leaves, Total Diet, Typical Diet, Oyster Tissue and Corn Bran, etc. The measured values for high boron concentration showed up to a 3% of the relative, but in a low concentration below 5 ppm, present values were higher than the certified ones.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.