Abstract

A reversed Delta III total shoulder prosthesis was retrieved post-mortem, eight months after implantation. A significant notch was evident at the inferior pole of the scapular neck which extended beyond the inferior fixation screw. This bone loss was associated with a corresponding, erosive defect of the polyethylene cup. Histological examination revealed a chronic foreign-body reaction in the joint capsule. There were, however, no histological signs of loosening of the glenoid base plate and the stability of the prosthetic articulation was only slightly reduced by the eroded rim of the cup.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.