Abstract

In this paper a measurements comparison between the National Institute of Metrological Research (INRIM), acting as pilot Laboratory, and nineteen Italian Secondary Calibration Laboratories in the field of low dc Electrical Resistance is presented and discussed. This comparison was carried out on commission by the Italian Accreditation Body of Calibration Laboratories ACCREDIA-DT. As traveling standards three Tinsley mod 3504C four terminal standard resistors to be calibrated in oil respectively at 5A and 20A (1mΩ), 1A and 5A (10mΩ), 0.1A and 1A (100mΩ) were involved. The temperatures at the measurements thermal regimen allowed a correction at 23°C of the results. The comparison results, shown graphically and as distribution of the normalized errors, were satisfactory, except in the case of a Laboratory whose measurements were outside its accreditation field.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.