Abstract

We determined transaminases in human blood serum with an amperometric glutamate biosensor. The probe was a hydrogen peroxide sensor assembled with appropriate selective membranes to enhance the probe specificity and lifetime. Calibration curves of glutamate were linear in the range 1-1000 mumol/L, with a response time of < 1 min. This probe was subsequently applied to the measurement of activities of aspartate and alanine aminotransferases in human sera. Analytical recovery studies demonstrated the suitability of the glutamate sensor by measuring 91-99% of added glutamate, 92-106% of added aspartate aminotransferase, and 101-105% of added alanine aminotransferase. Transaminase activity measured in 80 sera correlated well with results obtained with a spectrophotometric procedure.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.