Abstract

The purpose of this study is to analyze exposure to the time-varying magnetic field caused by worker movements in a 3-T clinical magnetic resonance imaging (MRI) scanner. Measurements of the static magnetic field (B) in the proximity of the MRI scanner were performed to create a detailed map of the spatial gradient of B, in order to indicate the areas at high risk of exposure. Moreover, a personal exposure recording system was used in order to analyze and compare exposure to the static magnetic field during different routine procedures in MRI. We found that for all of the performed work activities, exposure was compliant with International Commission on Non-Ionizing Radiation Protection levels. However, our findings confirm that there is great variability of exposure between different workers and suggest the importance of performing personal exposure measurements and of detailed knowledge of the magnetic field spatial distribution.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.