Abstract

( Quantitative analysis by Shpol'skii spectrofluorimetry of polynuclear aromatic hydrocarbons at 10 K with the mercury excitation line scattered by the matrix used as a standard intensity.) High-resolution Shpol'skii spectrometry in a matrix is particularly suitable for the determination of trace polynuclear aromatic hydrocarbons in complex mixtures. In the usual technique, the method of standard additions is backed up by internal calibration, which is tedious. A direct calibration technique is proposed in which the mercury excitation line, scattered by the polycrystalline matrix, serves as the standard intensity. Detection limits vary from 0.3 ng ml −1 for benzo(a)pyrene to 48 ng ml −1 for pyrene, with linear response over about 2 orders of magnitude. The dynamics and a statistical analysis of the confidence interval for the results are discussed.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.