Abstract

This study compared angular cephalometric values from scanned traditional lateral cephalograms versus CBCT generated slices. Traditional lateral cephalograms (using Orthoceph OC100) and volumetric CBCT imaging (using NewTom 3G Volume Scanner) were taken from 36 patients. Lateral cephalometric analysis using Dolphin 3D was performed on the respective imported radiographic data and ten angular measurements were evaluated. The reliability of measurements was evaluated with an intra-class correlation coefficient (ICC). Paired MANOVA was used to compare differences in measurements between the two image modalities. Inter-rater reliability demonstrated high reliability in 90% of the measurements (ICC >0.8). A high intra-rater reproducibility (ICC >0.8) was demonstrated in all angles. No significant difference in the angular measurements was detected between the two image modalities (paired MANOVA p >0.1, α =0.05). A lateral slice from CBCT imaging can be used to perform traditional cephalometric analysis with levels of precision and accuracy comparable to traditional lateral cephalometrics.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.