Abstract
Routine network analyzer calibration procedures in measurements of the dielectric properties of materials using an open-ended coaxial line probe are frequently inadequate and limit the accuracy of measurements. A calibration method, which makes use of liquids whose properties are well known, is proposed to alleviate this limitation. It is shown that even one liquid used as a standard in place of a matched load greatly improves the accuracy of measurements. Theoretical relationships and experimental results as well as some practical suggestions related to the application of this method are given.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: IEEE Transactions on Instrumentation and Measurement
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.