Abstract

We have developed a prototype of a new type of hybrid X-ray detector. It contains a thin wall (few μm) edge- illuminated lead glass capillary plate (acting as a converter of X-rays photons to primary electrons) combined with a microgap parallel-plate avalanche chamber operating in various gas mixtures at 1 atm. The operation of these converters was studied in a wide range of X-ray energies (from 6 to 60 keV) at incident angles varying from 0° to 90°. The detection efficiency, depending on the geometry, photon's energy, incident angle and the mode of operation, was between a few and ∼40%. The position resolution achieved was ∼50 μm in digital form and was practically independent of the photon's energy or gas mixture. The developed detector may open new possibilities for medical imaging, for example in mammography, portal imaging, radiography (including security devices), crystallography and many other applications.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.