Abstract
The sliding hip screw device (SHSD) is a commonly used implant to treat extracapsular neck of femur fractures. The most common described mode of failure is femoral head screw cut out. Mechanical failure is rare and when it occurs normally involves the plate rather that the sliding hip screw. Failure of the sliding hip screw itself is the most uncommon mode of failure and few cases have been reported in the literature. The mode of mechanical failure is usually multiple-cycle, low stress fatigue failure rather than low-cycle, high stress failure as a result of nonunion of the fracture. This case demonstrates a low cycle, high stress failure of the SHS and some of the potential causes for this are discussed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.