Abstract

There is a clinical need for improved ambulatory, frequent and unobtrusive monitoring of blood pressure and cardiac parameters like systolic time intervals. Truly unobtrusive wearable devices combining impedance cardiography with other sensors may be one possible solution. The IsenseU-BP+ device presented in this article measures single channel ECG, impedance cardiography and photo plethysmography at the chest. The device also measures activity and posture, as well as skin temperature. In this study, we report on the possibility to use these signals to measure pulse transit time for estimating blood pressure changes. Six subjects has been tested. Four of them showed good correlation between PTT and mean arterial pressure while two of the subjects had too low signal to noise ratio in the photoplethysmography signal for good estimation of PTT. Thus these results show that the quality of the raw data is promising for calculating a pulse transit time that shows good coherence with mean arterial pressure.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.