Abstract
We noticed an unprecedented sensing error: after a correctly sensed ventricular extrasystole followed by a correctly timed atrial stimulus, a ventricular impulse occurs that is not sensed, then the delivery of a ventricular stimulus can be observed at the set AV delay (225 ms) (Fig. 1). The subsequent stimuli appear to be sensed correctly, judging from the lack of any atrial or ventricular spikes. Fig. 1 Ventricular sensing failure, when the QRS is narrow IEGM scans eventually revealed the cause of the unexpected sensing error: while the R-wave amplitude exceeded 15 mV with a bundle branch block, we sensed a significantly lower R-wave amplitude with a narrow QRS (5.8–6.3 mV in the last two QRS) (Fig. 2). At a sensitivity of 5 mV, the pacemaker only intermittently sensed a narrow QRS. We have now lowered the ventricular sensitivity to an appropriate level. Fig. 2 Bundle branch dependent R wave amplitude
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.