Abstract

We present the case of a 24 year-old patient that developed massive hemolysis shortly after percutaneous left ventricular assist device, Impella 2.5, was placed. The hemolysis occurred without device alarm while the device was in the correct position. Further investigation of the device revealed fiber wrapped around the tip of the device, as a culprit. This case emphasizes on the special caution applied during device preparation to minimize the possible adverse events.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.