This paper focuses on the use of the spreadsheet technique to set up the uncertainty budget for the k0-standardisation method in Neutron Activation Analysis. The adopted measurement model included most of presently recognized error sources and was written to limit the covariances between input quantities. The calculations were implemented in a worksheet file and tested in a multi-elemental analysis of a biological material. Besides, it was demonstrated that the k0-standardisation turns to the relative-standardisation when the monitor element corresponds to the analyte element. The developed worksheet is available and suitable for the analysis of other materials in different experimental conditions.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.