Abstract

Prostate brachytherapy using radioactive seeds has emerged as an efficacious treatment option for men with prostate cancer. Due to the currently limited accuracy of delivering seeds, its scope has been restricted to primarily treating the entire prostate gland. In this paper we present a semi-automated seed delivery and tracking scheme for precise seed placement, to allow for more conformal treatment of localized prostate cancer. Standard seed-carrying needles are connected to a hand-held apparatus that the surgeon holds to steer the needle automatically as the surgeon manually inserts it. The apparatus is equipped with force sensors that are used to estimate the tissue properties during insertion. A needle steering controller employs ultrasound images of the needle in tissue to calculate the optimal steering manoeuvres online. A method is derived to track the position of the implanted seeds after the needle is withdrawn, allowing the surgeon to monitor the implant quality online. Experimental results over 90 seed implants indicate an average targeting accuracy of 0.46 mm.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.