Abstract
Heart failure (HF) is the number one cause of death in the world. B-type natriuretic peptide (BNP) is a recognized biomarker for HF and can be used for early detection. Field effect transistor (FET) biosensors have the ability to sense BNP in much shorter times than conventional clinical studies. The lowest limit of detection (LOD) of state-of-the-art HF FET biosensors is 100 fM and detection ranges are short, being less than 4 orders of magnitude. In this work, a B/N co-doped graphene oxide (GO) gel (BN-GO) was used as the channel material in an FET biosensor targeting BNP. The sensor was able to sense BNP in as little as 2 min, with an LOD as low as 10 aM and a wide linear detection range of 10 aM–1 μM, stretching over 11 orders of magnitude. The biosensor showed great selectivity and minimal response towards K+ and OH− ions and the human epidermal growth factor receptor (HER2) protein. This biosensor serves as a proof-of-concept of the ability of BN-GO gel FETs to be used for ultrasensitive biosensors.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.