Abstract

The development and validation of an ultra-high pressure liquid chromatography (UHPLC) tandem mass spectrometry (MS/MS) method was performed with the aim to be applied for the quantification of plasma teicoplanin concentrations in neonates. Pharmacokinetic data of teicoplanin in the neonatal population is very limited, therefore, a sensitive and reliable method for the determination of all isoforms of teicoplanin applied in a low volume of sample is of real importance. Teicoplanin main components were extracted by a simple acetonitrile precipitation step and analysed on a C18 chromatographic column by a triple quadrupole MS with electrospray ionization. The method provides quantitative data over a linear range of 25-6400ng/mL with LOD 8.5ng/mL and LOQ 25ng/mL for total teicoplanin. The method was applied in plasma samples from neonates to support pharmacokinetic data and proved to be a reliable and fast method for the quantification of teicoplanin concentration levels in plasma of infants during therapy in Intensive Care Unit.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.