Abstract

BackgroundThe cardiac troponin T (TnT) is a cardiospecific, highly sensitive marker for myocardial damage and is immediately released to bloodstream during the acute myocardial infarction (AMI). Surface plasmon resonance (SPR) immunosensor was developed for the quick detection of human cTnT in real time. MethodsAn SPR sensor streptavidin terminated self-assembled monolayer, which was used to binding biotinylated anti troponin T monoclonal antibodies, was developed. ResultsThe cTnT was determined from successive injections with a linear range from 0.03 up to 6.5 ng/ml. The detection limit was 0.01 ng/ml corresponding to a resonant angle change of 1.28 millidegrees. The system presented a good repeatability with 3.4% of variation between run after regeneration of the coated surface with a solution of 1% SDS. The effects of blocking of non-specific adsorption using different solutions were studied. It was possible to measure the cTnT without dilution of the human serum with good specificity and reproducibility. ConclusionThis sensor is practical and offer quick response in interval of 800 s.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.