Abstract
Continuous patient monitoring enables early detection and treatment of diseases. Small and wireless devices are more comfortable to the patient and simplify the implantation procedure. This paper presents the design and partial testing of an integrated RF-powered temperature sensor suitable for the human body temperature range, from 35 to 42oC. The device was designed using a 130-nm standard CMOS technology and it receives energy from a 900-MHz signal that has power as low as −10 dBm. A calibration method was designed to obtain conversion errors smaller than 0.2oC.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.