Abstract

A modified PVC-membrane electrode with tetra-butylammonium bromide — Clinoptilolite nano-particles (TBA-NCP) showed good Nernstian slope (29.9±0.6mV per decade of oxalate concentration) in concentration range of 3.1×10−7–8.3×10−1molL−1 with a detection limit of 1.5×10−7molL−1. The best performance was obtained with a membrane composition of 31.5% PVC, 62.5% DOP and 6% TBA-NCP in the temperature range of 20–35°C and the pH range of 4–9. The fast response time and good reproducibility over a period of 3months are other characteristics of the sensor. The proposed electrode was successfully used as an indicator electrode in titration of oxalate ions with CaCl2 solution. The proposed electrode was also used in direct potentiometric determination of oxalate in many real samples such as: mushroom, black and green tea, spinach and beet.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.