Abstract

To develop a rapid and reliable method, using an octopole reaction system (ORS) ICP-MS, capable of monitoring trace levels of Co and Cr in whole blood samples from hip arthroplasty patients with metal-on-metal prostheses. Whole blood is diluted 10-fold with an alkaline diluent and analyzed using an Agilent 7500 CE ORS-ICP-MS. Limit of quantification of 0.03 μg/L Co and 0.20 μg/L Cr in patient samples. <6% covariance obtained for quality control materials analyzed over 10 runs. This method is capable of monitoring trace levels of Co and Cr in diluted whole blood samples with a vial to vial run time of approximately 2 min. Results are comparable to those obtained using high resolution (HR) ICP-MS with sample digestion.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.