Abstract

The principle of a new tomograph (TOHR) dedicated for small volume analysis with very high resolution is presented. The authors use uncorrelated multi-photon (X- or gamma rays) radioisotopes and a large solid angle focusing collimator to produce tomographic images without a reconstruction algorithm. With this device, detection efficiency and resolution are independent and submillimeter resolution can be achieved. A feasibility study shows that collimators made with a stack of chemically etched plates can achieve the predicted performances of TOHR. The authors discuss its potential in rat brain tomography by simulating a realistic neuropharmacological experiment using a 1.4 mm resolution prototype of the TOHR under development.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.